The PAMAS SBSS is a particle counting system for batch sampling. The PAMAS SBSS is the standard instrument used as a reference in many testing labs. The instrument may be programmed to measure a sample without the need of an operator, like pre-filling the internal dead volume, degassing the sample with vacuum, pressurising the sample for a number of measurement runs, and relieving pressure at the end. High accuracy and repeatability, fast purging and back flush possibility are the key benefits of the system.
Applications:
The system is used for hydraulic oils, insulation- and turbine oils and organic fluids or solvents. With the software PAMAS USP, the system may be also used for pharmaceutical applications with higher viscosities.
Viscosity and pressure:
max. 5 bar - 600 cSt with standard accessory pump
max. 10 bar - 1500 cSt with high pressure pump or external pressure supply
Volumetric particle sensors:
PAMAS HCB-LD-50/50
Size range:
1 - 400 μm with calibration according to ISO 21501
1 - 100 μm with calibration according to ISO 4402
4 - 70 μm(c) with calibration according to ISO 11171
Max. particle concentration: 24,000 p/ml at 25 ml/min
PAMAS HCB-LD-25/25
Size range:
1 - 200 μm with calibration according to ISO 21501
1 - 100 μm with calibration according to ISO 4402
4 - 70 μm(c) with calibration according to ISO 11171
Max. particle concentration: 120,000 p/ml at 10 ml/min
Software:
PAMAS USP for pharmaceutical applications (21 CFR Part 11 compliant)
PAMAS PMA for the contamination control of clean liquids
Option:
PAMAS SBSS-HP
high pressure version
max. 30 bar - 20.000 cSt with external pressure supply